Load cell cable is not a passive connection. It carries the excitation current to the bridge and the millivolt signal back, and every metre of it has resistance. On a short run the effect is negligible; on a long one it is a calibration error that nobody recorded.
Where the error comes from
A strain gauge bridge is a resistive divider. Its output is proportional to the excitation voltage across it. If the excitation voltage at the cell is lower than the instrument assumes, the output per unit load is lower, and the instrument under-reads.
The cable adds resistance in two places: the excitation pair, which carries the bridge current, and the signal pair, which carries almost none. Only the first matters for the excitation drop; the second matters only for noise pickup.
With a 350 ohm bridge drawing about 30 mA from a 10 V supply, each ohm of loop resistance in the excitation path drops 30 mV. Five ohms of loop resistance — which is roughly 100 metres of small cable — costs 150 mV, or 1.5 % of the excitation. That is 1.5 % of the reading, present at every point of the range.
| Connection | What it compensates | When it is needed |
|---|---|---|
| 4-wire | nothing — excitation is sensed at the instrument | short cable, stable temperature |
| 6-wire | cable drop, by sensing at the cell | long cable, or cable that changes with temperature |
Why 6-wire exists
A 6-wire cell brings two extra conductors back from the bridge so the instrument can measure the actual excitation voltage at the cell and correct for the drop. The correction is continuous, so it also handles the fact that copper resistance rises with temperature — a long cable run in a hot plant room changes resistance by several percent between winter and summer.
This is the reason a 4-wire installation that calibrates perfectly in the workshop drifts after installation in a long run: the calibration absorbed the drop at one temperature, and the drop changed.
The practical limits
As a rough guide, 4-wire is acceptable up to about 10 metres with standard cable, and 6-wire beyond that. The number is not a standard; it depends on the bridge resistance, the excitation voltage and the accuracy required.
The calculation is simple enough to do before installation: loop resistance of the excitation pair, times the bridge current, gives the drop. Compare that with the excitation voltage and decide whether the resulting error is acceptable for the application.
What else cable length affects
Noise. The signal is a few millivolts. A long run picks up more interference, and the shielding matters more than the length. Shield connected at one end, usually the instrument end.
Capacitance. Long cable adds capacitance across the signal pair, which slows the response and can make a fast-reading instrument unstable. This shows up as a reading that hunts rather than one that is offset.
Temperature gradients. A cable run that passes from a cold store to a warm room creates thermoelectric voltages at the junctions if the connections are at different temperatures. Keeping both ends of a junction at the same temperature is cheaper than compensating for it.
What to record
The installation record should carry the cable type, the length, the connection scheme (4 or 6 wire) and the excitation voltage actually measured at the cell. That last value is the one that explains a reading discrepancy six months later, and it is the one most often missing.
If a system is recalibrated after a cable change, the record should note the change. A calibration that does not mention the cable looks valid but describes a different system.