Four pressure numbers appear on a pressure sensor datasheet and they are routinely mixed up in specifications, in purchase orders and in failure investigations. The definitions follow the transmitter terminology of IEC 60770-1, where rated range, overload and ultimate pressure are distinguished, and using the wrong one as a selection criterion is how a project ends up with a sensor that reads correctly for a month and then drifts permanently.
The four limits
| Rated capacity | The pressure at which accuracy is specified. Above it, the accuracy figures no longer apply even if the sensor survives. |
| Proof pressure | A pressure the sensor can withstand without permanent change to its characteristics. Typically 1.5 to 2 times rated capacity. |
| Burst pressure | A pressure the sensor can withstand without rupturing. Typically 3 to 5 times rated capacity, and it says nothing about function afterwards. |
| Overpressure limit | Sometimes used as a synonym for proof pressure, sometimes for a higher non-destructive limit. Always read the definition, never the word. |
The distinction that matters in selection: proof pressure is a functional promise, burst pressure is a containment promise. A sensor taken to burst pressure may still hold pressure and be useless as a sensor.
What "without permanent change" means in practice
Testing proof pressure means applying it, holding it briefly, releasing it, and then re-checking the sensor at rated capacity. The acceptance criterion is that zero and span have not moved beyond the datasheet tolerance. A sensor whose zero shifts by 0.5 percent of full scale after a proof test has failed the proof test even though it still works.
This is why proof pressure is the number to use when a system can see a transient above the working range — water hammer, a pump start, a valve closure. The transient does not need to burst anything to cause a problem; it only needs to shift the zero.
The over-range failure that looks like drift
A pressure sensor that has been over-ranged usually does not fail outright. It develops a permanent offset. The field symptom is a reading that is consistently high or low by a fixed amount, with normal sensitivity to pressure change. Three causes, in order:
- Diaphragm yield. The sensing diaphragm is deformed past its elastic limit and no longer returns to the same shape. A 10 bar sensor taken to 15 bar is 50 percent over range, which is inside many proof limits and still enough to move the zero. The offset is permanent and the sensor must be replaced.
- Bond degradation. The strain gauge bond or the dielectric layer is damaged, which changes the bridge balance.
- Seal extrusion. An O-ring or a welded seam deforms, changing the internal volume and the preload on the diaphragm.
Distinguishing an over-range offset from electronic drift is straightforward: apply zero pressure and check the output. Electronic drift usually moves slowly and is temperature-correlated; over-range offset is fixed and appears after a specific event.
Static versus cyclic pressure
Rated capacity assumes a static or slowly varying load. A sensor in a pressure pulsation application sees millions of cycles, and fatigue becomes the limiting mechanism rather than proof pressure. The relevant datasheet figure is then the cycle life at a stated pulsation amplitude, not the proof pressure.
Two rules for pulsating service:
- Derate. A sensor rated 10 bar used at 10 bar in a pulsating line will have a shorter life than the same sensor used at 7 bar.
- Check the mounting torque and the seal type. A pulsating load works a threaded seal loose, and a loose seal leaks before the sensor fails.
Reading the datasheet as a specification
When writing a purchase specification, three lines do more work than the rest:
- Rated capacity and the accuracy at it, with the temperature band over which the accuracy applies. Accuracy quoted at 25 degrees C only is not an accuracy for a plant room.
- Proof pressure, stated as a multiple of rated capacity, so the sensor survives the transients the process actually produces.
- Long-term stability, usually percent of full scale per year. This is the figure that decides the calibration interval.
A fourth line is worth asking for even when it is not on the datasheet: the burst pressure with the seal type fitted. Burst pressure of the sensing element and burst pressure of the process connection are different numbers, and the lower one is the one that matters.
The pressure sensor range lists rated capacity, proof and burst ratios for each series; the product index links the process connections and seals used with them.