Eccentricity is the quietest of the load cell tests and the one most often skipped. The cell is loaded at the centre, the reading is right, and the scale ships. In the field the same scale is loaded at the edge, and the error appears.
The test position is specified as one third of the platform length from the centre, because that is where the combination of moment arm and structure stiffness produces the largest error on a typical platform.
What the load does to the cell
| Load position | Effect on the cell | Typical error |
| Centre | Pure vertical force | Reference reading |
| 1/3 position | Vertical force plus a moment | Largest error on most platforms |
| Corner | Moment plus platform twist | Error depends on the frame, can be larger |
| Edge, long side | Moment about the short axis | Intermediate |
| Off-platform | Not a valid test | Damage risk |
The tolerance is usually expressed as a fraction of the verification scale interval, and for a 3000 division instrument it is tight enough that a flexible platform will fail even with a perfect cell.
Separating cell error from frame error
- Test the cell alone. Load it with a fixture that applies the same moment, without the platform.
- Then test with the platform. The difference is the frame.
- Check the fixings. One bolt tighter than the others tilts the cell and produces a repeatable eccentric error.
- Check for a warped base. A base that rocks under load produces an error that changes with load position and with time.
Instrumentation that makes the test quick
Doing the test properly needs four positions and one reference weight, moved in a fixed order, with a reading at each. A weighing instrument with a print key and a stable-reading indicator turns this from twenty minutes of note-taking into four button presses, and the record is the printout.
Our weighing load cells are tested for eccentricity at the 1/3 position on a certified platform, and the technical notes cover the fixture design that separates cell error from platform error.
Standards referenced: OIML R76-1 clause 5.2 (eccentricity), OIML R60, GB/T 7723 for fixed platform scales.
The four positions, in the order that finds faults
Load the centre, then the 1/3 position on the long axis, then the 1/3 position on the short axis, then a corner. Four readings against one reference weight, about 15 minutes on a 300 kg platform. A corner reading that is 2 divisions low while the 1/3 readings are fine points at the frame, not the cell.
Repeat at half load as well: eccentricity errors often scale differently from linearity errors.
On a 300 kg platform with 3,000 divisions, one division is 0.1 kg. A healthy platform reads within 2 divisions at the 1/3 position and within 3 at the corner; a reading 8 divisions low at the corner points at a fixing torque of about 20 N·m on one bolt against 45 N·m on the others, not at the cell.