A load cell produces about 2 mV per volt of excitation. Feed it 5 V through 20 m of cable and the cell may see 4.95 V; the instrument assumes 5 V and the reading is one percent low. Nothing is broken, and the calibration certificate is still valid.
Six-wire connection exists to remove that error: two extra conductors carry no current and tell the instrument what the bridge actually received.
The arithmetic of a long cable
Standard cable is about 0.06 Ω per metre per conductor. A 20 m run has 2.4 Ω in each of the two excitation conductors, so 4.8 Ω in the loop. Against a 350 Ω bridge that is 1.4 % of the excitation lost, and therefore about 1.4 % of the reading lost.
| Bridge | Cable loop resistance | Excitation loss | Reading error |
| 350 Ω | 1 Ω | 0.3 % | 0.3 % low |
| 350 Ω | 5 Ω | 1.4 % | 1.4 % low |
| 350 Ω | 10 Ω | 2.8 % | 2.8 % low |
| 1000 Ω | 5 Ω | 0.5 % | 0.5 % low |
| 350 Ω, six-wire | Any | Corrected by the sense loop | No error from cable |
Two percent on a 300 kg scale is 6 kg. That is the difference between a compliant weighbridge and a rejected one, and it comes from the cable.
When six-wire is worth the extra conductors
- Cable longer than about 10 m. Below that the error is usually inside the instrument tolerance.
- Multiple cells in parallel. The junction box adds resistance that a four-wire system folds into the gain.
- Installations where the cable may be replaced. A different cable length then changes nothing.
- Regulated excitation supplies. A six-wire instrument corrects at the bridge, not at the terminals.
The mistakes that cancel the benefit
A six-wire cell wired as four-wire, with the sense leads cut and taped, behaves exactly like a four-wire cell and nobody notices until the cable is replaced. A junction box that ties the sense leads to the excitation at the box, rather than at the instrument, also removes the correction for the run back to the instrument.
Check the wiring by measuring the voltage at the cell while the instrument reads the sense line: the two should agree within a millivolt. Our load cell and junction box range includes six-wire versions with the sense loop brought out to the connector, and the technical notes give the cable resistance limits for parallel installations.
Standards referenced: OIML R76 (non-automatic weighing instruments), OIML R60, IEC 61158 for the industrial cabling practice.
The 5 minute test that tells you which one you have
Measure the excitation voltage at the instrument and at the cell with the system powered. A difference of 30 mV on a 5 V supply is 0.6 percent of the reading; if the sense leads are connected, the instrument corrects for it and the reading will not move when you add 10 m of cable.
- Four-wire: the reading drops when the cable gets longer.
- Six-wire: the reading stays put.
- Six-wire wired as four: the reading drops, and nobody notices until the cable is replaced.