Load cell data sheets list two overload figures, and reading them as one number is how a cell gets blamed for a fault it did not have. Safe overload is the load the cell can take repeatedly without changing; ultimate overload is the load it can survive once. Between them lies a zone where the cell still works but no longer measures what it did.
The two ratings
| Rating | Typical value | What it means |
|---|---|---|
| safe overload | 120 to 150 % | repeatable, no change to the cell |
| ultimate overload | 200 to 400 % | survives, may be permanently shifted |
A cell rated 300 % ultimate can absorb a 150 % event — a dropped load, a truck driven onto a platform — and come out working. What it may not do is come out measuring correctly.
What happens inside during an overload
Three things, in order of how easily they are detected:
The strain gauge bond. The adhesive between foil and steel is loaded beyond its elastic range. It creeps, and the gauge's position changes slightly. The result is a zero offset that does not return.
The spring element. A small permanent set in the metal, typically a few micrometres at the strain level involved. This changes the sensitivity — the cell now produces slightly fewer millivolts per kilogram.
The mechanical stops. Many cells have overload stops that contact before the element yields. If the stop is hit, the load path changes and the cell's response becomes non-linear near capacity.
Why it looks like a calibration problem
A zero shift is the classic symptom. The instrument reads 12 kg with nothing on the platform, the operator re-zeros, and the scale appears fine. What has actually happened is that the cell's zero moved and the instrument's zero was moved to compensate.
The error then appears at the top of the range: with the zero corrected, the span is now wrong, because the cell's sensitivity changed too. A check weight near capacity reads low, and the assumption is that the instrument needs calibrating.
Recalibrating restores the reading at the check weight and leaves the middle of the range wrong. The scale passes its calibration check and continues to under-read at intermediate loads.
How to tell overload from drift
Three questions separate the two:
Did the zero move by more than the drift specification? Drift is a few hundredths of a percent per year. A shift of 0.5 % is not drift.
Did the span change at the same time? Zero and span moving together points at a physical event. Zero alone is more likely thermal or electrical.
Is there a witness? Overload stops that have been contacted show a bright mark or a flattened area. Inspect them before anything else.
Preventing it
Mechanical stops external to the cell. A stop that takes the load before the cell does, set a few percent below the safe overload, is cheaper than a higher-rated cell.
Correct capacity selection. A cell chosen at exactly the working load has no margin for the dynamic component. A conveyor's impact load can be several times the static weight.
Load-path inspection after any event. A dropped load or a vehicle on the platform is a reason to check the cell, not a reason to assume it survived because the reading looks plausible.
After an overload
A cell that has been overloaded and now shows a zero shift should be replaced, not recalibrated. Recalibration corrects the endpoint and leaves the middle wrong, and the error is invisible to a single-point check.
The exception is a cell with a documented, repeatable offset that has been corrected by the manufacturer's own procedure. Without that documentation, a shifted cell is a cell whose future measurements cannot be trusted.