A load cell specification lists a temperature coefficient, and reading it as a single number is how a scale gets calibrated in a warm room and drifts in a cold store. There are two coefficients, they behave differently, and they need different responses.
The two coefficients
Zero drift is a shift in the output at no load. The reading with an empty platform changes as the temperature changes. It is usually quoted as a percentage of full scale per 10 °C.
Span drift is a change in sensitivity. The output per kilogram changes, so a load near capacity reads differently even after the zero is corrected.
| Coefficient | Symptom | Corrective action |
|---|---|---|
| zero drift | empty platform reads non-zero | re-zero, or compensate in the instrument |
| span drift | check weight reads high or low | span calibration, cannot be zeroed out |
Why zeroing hides the problem
An operator who sees a drifting zero presses the zero button. That corrects the empty-platform reading and leaves the span error untouched. The scale now passes an empty check and under-reads or over-reads at every load above zero.
This is the most common way a temperature problem survives routine calibration: the zero is corrected daily, the span is checked yearly, and the error in between is never measured.
Where the drift comes from
Two mechanisms, and they are additive:
The strain gauge itself. The gauge factor of a foil gauge changes with temperature. This affects span.
The bridge circuit. The bridge completion resistors and the excitation circuit have their own temperature behaviour, which affects zero. Careful designs use matched resistors and, on better cells, a compensation network built into the cell.
A cell with good compensation still drifts; it drifts less. The specification tells you how much, and the number is per 10 °C, not per degree.
The installation makes it worse
Three installation effects add to the cell's own drift:
Temperature gradient across the cell. If one side of the cell is warmer than the other, the bridge sees a differential that looks like a shift. A cell in a draught from a chiller door will have a gradient that changes with the door.
Cable thermocouples. The junction between the cell's cable and the extension cable is a thermocouple if the two metals differ. A temperature change at that junction adds a voltage that the instrument reads as load.
Mechanical thermal expansion. The structure the cell is mounted in expands with temperature, changing the preload on the cell. This appears as zero drift but is not the cell's coefficient — it is the installation's.
How to tell them apart
Put the scale in a chamber and record the reading at no load and at a check weight across the temperature range. Then:
If the no-load reading moves and the check weight reading moves by the same amount, the error is zero drift.
If the check weight reading moves by more than the no-load reading, part of the error is span drift.
If the error depends on how fast the temperature changed, rather than on the temperature itself, it is a gradient effect, not a coefficient.
What to specify
A cell with a quoted coefficient of 0.02 % per 10 °C in a 30 °C range contributes 0.06 % of full scale — which on a 1000 kg scale is 600 g. Whether that matters depends on the application, and the number is worth calculating before choosing the cell rather than after the scale fails a check.
For applications where the temperature range is wide, a cell with active compensation or a four-wire connection that senses temperature is worth the cost. For a scale in a stable room, the standard cell is fine and the specification is there to be checked, not feared.
Recording it
The useful record is a table: temperature, no-load reading, check-weight reading. Three columns, taken at four or five temperatures. That table separates zero from span without any calculation, and it is the only evidence that distinguishes a cell problem from an installation problem.