A load cell cable has a shield around the signal pair. Where it is connected decides whether the shield does its job or becomes a source of error.
What the shield is for
The signal from a load cell is a few millivolts. Any electrical noise coupled onto the signal leads is comparable to the signal itself. The shield intercepts that noise and provides a path to ground so it does not reach the signal conductors.
For this to work, the shield must be connected to a reference potential — but only at one point.
Why both ends is wrong
When the shield is grounded at both the cell end and the instrument end, and the two grounds are at slightly different potentials (which they always are), a current flows through the shield.
That current has two effects:
It couples into the signal pair. The shield is capacitively and magnetically coupled to the conductors inside it. A current in the shield induces a voltage in the signal pair.
It creates a ground loop. The loop picks up magnetic fields from nearby cables and motors, and the induced voltage appears as a slowly varying offset.
| Shield connection | Noise pickup | Symptom |
|---|---|---|
| one end (instrument) | lowest | stable reading |
| both ends | ground loop current | drift that follows nearby equipment |
| neither end | no path to ground | reading jumps with static and radio |
Which end to ground
Convention is the instrument end, and it is the right convention for a reason: the instrument's ground is the reference the measurement is made against. Grounding there keeps the shield at the same potential as the measuring circuit.
Grounding at the cell end is used in some installations, particularly where the cell is in a well-grounded structure and the instrument is not. The choice should be documented, because the next person to service the cable will otherwise reconnect it by habit.
The visible symptom of a two-end ground
The scale reads correctly when nothing else is running. When a motor starts, or a welder operates, or a variable-speed drive ramps, the reading moves.
Then it settles. Then it moves again. The operator reports "the scale drifts" and the maintenance response is usually to re-calibrate — which does not change anything.
A quick test: disconnect the shield at the cell end and observe. If the drift stops, the ground loop was the cause.
Where the shield should not be connected
Not to the enclosure of a junction box that is itself earthed at a different point. Not to a cable tray. Not to the machine frame if the frame is a separate earth path.
The shield should be continuous from the cell to the instrument and connected at exactly one point along that path.
Practical details
Use a proper cable gland for shielded cable. The shield must be terminated with 360-degree contact, not with a pigtail. A pigtail creates an inductor that blocks high-frequency noise while passing low-frequency ground currents — the worst combination.
Keep the signal cable away from power cables. Separating them by 300 mm is more effective than any shield. Where they must cross, cross at right angles.
Check continuity of the shield. A shield broken in the middle and grounded at both ends behaves like neither.
What to record
Which end is grounded, the cable route, and whether a pigtail or a gland was used. Three items that let the next person reproduce the installation rather than guess.