A load cell datasheet lists creep as a single number, usually around 0.02 percent of full scale over 30 minutes. It looks like a minor specification. On a process scale that holds a load while a tank fills, it is the drift that the operator compensates for by hand without knowing why.
What creep is
Apply a load, hold it steady, and watch the reading. It does not stay put: it changes slowly in the direction of the load, then stabilises. That change is creep, and it is a property of the strain gauge installation, not of the steel.
The gauge foil and the adhesive between foil and substrate continue to move for a period after the load is applied. The movement is elastic but time-dependent, and it is why readings taken immediately after loading differ from readings taken a minute later.
| Time after loading | Typical behaviour | On a 30 t scale |
|---|---|---|
| 0 to 10 s | settling, mechanical and electrical | tens of kilograms of apparent change |
| 10 s to 30 min | creep | about 6 kg at 0.02 % FS |
| after 30 min | stable, or still drifting if overloaded | a continuing drift is a defect |
Why 30 minutes
The test duration is a convention that captures most of the creep for most cells. It does not mean the cell stops creeping at 30 minutes; it means the remaining change is small enough to be ignored for the accuracy class.
For a process that holds a load for eight hours, the remaining creep is not negligible, and a cell that meets the 30-minute specification can still drift visibly over a shift.
Creep and creep recovery
Two numbers, often confused. Creep is the change under a held load. Creep recovery is the residual offset after the load is removed and the cell is allowed to return.
A cell with good creep but poor recovery returns close to zero slowly, which on a batch scale accumulates into the tare. A cell with poor creep but good recovery drifts during the hold and then returns cleanly. The first is a problem for weighing; the second is a problem for process control.
What causes excessive creep
Three sources:
The adhesive layer. Too thick an adhesive layer allows relative movement between foil and substrate. It creeps more, and it creeps for longer.
Incomplete cure. Adhesive that has not fully cured continues to flow under load. The symptom is a cell whose creep improves over the first months of service, which is the opposite of what wear would do.
Mechanical creep in the load path. A bolted joint, a pin in a clearance hole, a mounting surface that is not flat. The cell is blamed for movement that happens outside it.
Separating cell creep from installation creep
Load the cell directly, with the load applied through a fixture that bypasses the installation, and repeat the hold. If the creep disappears, the cell is fine and the installation is the source.
This is a thirty-minute test that settles an argument which otherwise runs for weeks, and it is rarely done because it requires dismantling something.
Compensating for it in the instrument
Some instruments offer creep compensation: a time-based correction applied to the reading. It works for a known cell type and a known load history, and it fails when either changes.
The honest use of such a feature is to correct a known, characterised creep curve for a specific cell model in a specific application. Turning it on because it exists is how a scale ends up over-correcting at low loads.
What the operator sees
A scale that reads slightly high immediately after loading and settles down over a minute. The operator learns to wait, or to press the hold button at what feels like the right moment.
Both behaviours hide the same thing: the reading at t=0 and the reading at t=60 s differ by more than the resolution the process needs. That is worth measuring, because it sets the minimum settling time the process must allow.
What to record
Reading at 1 s, 10 s, 60 s, 5 min and 30 min after applying a known load, and the same sequence after removing it. Five points on each curve.
Two curves plotted together show creep, recovery and settling in one picture, and they are the only evidence that distinguishes a cell problem from an installation problem.