A weighbridge reads true through the day and shows a few kilograms of offset at six in the morning. Nobody changed anything. What changed is the temperature, and the cell's temperature coefficient — a number on the datasheet that almost nobody reads — has been reporting it all along.
The two coefficients
A load cell datasheet lists two temperature effects, and they behave differently:
Temperature effect on zero (often written TC zero, or "zero balance"). The reading with no load changes with temperature. It is an offset, and it shifts the whole calibration curve up or down.
Temperature effect on span (TC span, or "sensitivity"). The reading with a load changes proportionally — a 1000 kg load reads differently at 10 °C and at 30 °C. It is a gain error, and its effect is zero at zero load.
Both are quoted as a percentage of full scale per 10 K, typically around 0.02 %/10 K for a good industrial cell.
| Effect | What it does | On a 30 t scale at 0.02 %/10 K |
|---|---|---|
| zero | offsets the curve | 6 kg per 10 K |
| span | scales the curve | 6 kg at full load per 10 K |
Why the drift appears at night
The distinction between zero and span explains the pattern. At night the scale is empty, so a span error has nothing to act on — the reading should be zero. If the empty scale shows an offset, it is the zero coefficient, or something worse than a coefficient.
The usual causes, in order of likelihood:
Genuine zero coefficient. The cell's bridge is not perfectly compensated. A 15 K overnight swing on a cell with 0.02 %/10 K moves a 30 t scale by about 9 kg.
Apparent strain from the structure. The cell is bolted between a frame and a platform, and those expand at different rates. The cell is being strained by the installation, not by the load. This is not on the datasheet because it is not the cell's property.
Cable and connector effects. A junction box that cools draws moisture, or a connector's contact resistance changes. Small in microvolts, but the signal is only a few millivolts.
Instrument zero drift. The indicator's own amplifier has a temperature coefficient, and it is sometimes worse than the cell's.
Separating them
A short procedure separates the four:
1. Load the scale with a known mass and watch through a temperature change. If the reading at load changes by the same absolute amount as the zero reading, the cause is zero drift. If it changes proportionally, it is span.
2. Repeat with the cell at a stable temperature but covered, and the structure exposed. If the drift disappears, the cause is structural.
3. Disconnect the cable at the indicator and substitute a stable millivolt source. If the indicator reading still drifts, the indicator is responsible.
This takes an afternoon and settles a question that otherwise gets argued for years.
Compensation is not removal
Cells are compensated with thermistors in the bridge circuit, and good compensation reduces the coefficient by an order of magnitude. It does not remove it, and the residual is what the datasheet number describes.
Two practical consequences: a scale used across a wide temperature range needs either a wider tolerance, or a live temperature input so the instrument can apply a correction. Many industrial indicators accept a temperature sensor for exactly this reason.
What to do about it operationally
Zero before each weighing session. A tare is not a nuisance step; it removes the accumulated zero error at the moment it matters.
Record the temperature at calibration. A calibration certificate without the ambient temperature cannot be compared with the next one.
Calibrate at a representative temperature. A scale calibrated in an air-conditioned workshop and installed in an unheated bay is calibrated at the wrong point of its range.
The structural contribution is usually the biggest
In field investigations, the structural term is more often the dominant one than the cell's coefficient, and it is the one nobody computes. A steel platform on a concrete base, a stainless cell in a mild steel frame, a load path with a bolted joint that changes preload — each produces apparent strain that looks exactly like a cell defect.
The test is simple: put a dummy cell, or a shorting link, in place of the real cell and repeat the temperature cycle. If the indicator still moves, the movement is in the structure or the instrument, and the cell is being blamed for something it did not do.
What to record
Zero reading and temperature at the start and end of a shift, the load reading at a check weight at both times, and the ambient range over 24 hours. Four numbers that identify which coefficient is acting, and whether the answer is compensation, installation, or acceptance of a wider tolerance.